RESmart Auto CPAP System

   
      Thumb Thumb Thumb Thumb

G2S A20

  • Device Size

    - Dimensions: 274 mm × 184 mm × 115 mm.

    - Weight: 1.9 kg

    - Water capacity: To maximum fill line 360 mL

  • Product Use, Transport and Storage
    Operation

    - Temperature: 5℃ to 35℃ (41℉ to 95℉).

    - Humidity: 15% to 93% Non-condensing

    - Atmospheric Pressure: 760 to 1060 hPa

  • Product Use, Transport and Storage
    Transport and Storage

    - -25℃ to 70℃ (-13℉ to 158℉)

    - 15% to 93% Non-condensing

    - 760 to 1060 hPa

  • Heated Humidifier

    Humidifier Settings: off, 1 to 5 (95℉to 154.4℉ / 35℃ to 68℃) Humidifier Output: No less than 10 mg H2O/L Environmental Conditions: Maximum airflow, 35℃, 15% relative humidity

  • Mode of Operation

    Continuous

  • Work Mode

    Auto CPAP, CPAP

  • SD Card

    The SD card can record patient data and fault information

  • AC Power Consumption

    100 - 240 V AC, 50/60 Hz, 2.0 A

  • Type of Protection Against Electric Shock

    Class II Equipment

         

Specifications

 
  • Degree of Protection Against Electric Shock

    Type BF Applied Part

  • Degree of Protection Against Ingress of Water

    IP22

  • Pressure Range

    4 to 20 hPa (in 0.5 hPa increments), ≤30 hPa under single fault conditions.

  • Static Pressure Stability

    ± 0.5 hPa

  • Ramp

    The ramp time ranges from 0 to 60 minutes

  • Sound Pressure Level

    < 28 dB, when the device is working at the pressure of 10 hPa

  • Sound Power Level

    < 38 dB, when the device is working at the pressure of 10 hPa

  • Maximum Flow

    * Test Pressures (hPa) : 4 10 15 20

    * Measured Pressure at the Patient Connection Port (hPa) :

    - 3 9 14 19

    * Average Flow at the Patient Connection Port (L/min) :

    - 85 135 140 140

  • Automatic leak compensation

    The pressure error ≤ ± 1 hPa when the air leakage is no more than 70 L/min.

 
  • Pressure

    - Range: 0 to 20 hPa

    - Margin of Error: ± (0.4 hPa + 4%)

  • SpO2

    - Range: 35% to 100%

    - The margin of error for SpO2 between 70% and 100% is ± 3%.

    - No strict accuracy requirements for SpO2 below 70%.

  • Pulse Rate

    - Range: 30 to 240 BPM

    - Margin of Error: ±2%

  • Wavelengths

    - Red: 663 nanometers

    - Infrared: 890 nanometers

  • Maximal Optical Output Power

    Less than 1.5 mw maximum average.

  • Tube

    Length: 6 ft. (1.83 m)

  • Maximum Delivered Gas Temperature

    ≤ 43℃

  • The Form and the Dimensions of the Patient Connection Port

    The 22 mm conical air outlet complies with ISO 5356-1.