- Dimensions: 274 mm × 184 mm × 115 mm.
- Weight: 1.9 kg
- Water capacity: To maximum fill line 360 mL
- Temperature: 5℃ to 35℃ (41℉ to 95℉).
- Humidity: 15% to 93% Non-condensing
- Atmospheric Pressure: 760 to 1060 hPa
- -25℃ to 70℃ (-13℉ to 158℉)
- 15% to 93% Non-condensing
- 760 to 1060 hPa
Humidifier Settings: off, 1 to 5 (95℉to 154.4℉ / 35℃ to 68℃) Humidifier Output: No less than 10 mg H2O/L Environmental Conditions: Maximum airflow, 35℃, 15% relative humidity
Continuous
Auto CPAP, CPAP
The SD card can record patient data and fault information
100 - 240 V AC, 50/60 Hz, 2.0 A
Class II Equipment
Type BF Applied Part
IP22
4 to 20 hPa (in 0.5 hPa increments), ≤30 hPa under single fault conditions.
± 0.5 hPa
The ramp time ranges from 0 to 60 minutes
< 28 dB, when the device is working at the pressure of 10 hPa
< 38 dB, when the device is working at the pressure of 10 hPa
* Test Pressures (hPa) : 4 10 15 20
* Measured Pressure at the Patient Connection Port (hPa) :
- 3 9 14 19
* Average Flow at the Patient Connection Port (L/min) :
- 85 135 140 140
The pressure error ≤ ± 1 hPa when the air leakage is no more than 70 L/min.
- Range: 0 to 20 hPa
- Margin of Error: ± (0.4 hPa + 4%)
- Range: 35% to 100%
- The margin of error for SpO2 between 70% and 100% is ± 3%.
- No strict accuracy requirements for SpO2 below 70%.
- Range: 30 to 240 BPM
- Margin of Error: ±2%
- Red: 663 nanometers
- Infrared: 890 nanometers
Less than 1.5 mw maximum average.
Length: 6 ft. (1.83 m)
≤ 43℃
The 22 mm conical air outlet complies with ISO 5356-1.